Abstract
The stripping voltammetry of Tellurium (IV) in 0.1M HC10, using gold electrodes is described. Detection limits for solid gold, in situ gold-plated and externally gold-plated rotating glassy carbon disk electrodes are presented. No significant increase in sensitivity is reported for the use of solid gold. electrodes, and 0.25 ppm Tellurium (IV) can be determined by anodic stripping voltammetry at an in situ gold-plated rotating glassy carbon disk electrode. The determination of tellurium in National Bureau of Standards, Standard Reference Material 1623a (Trace Elements in Coal) is presented.
Recommended Citation
Posey, Robert Steve, "Stripping Voltammetry of Tellurium (IV) In 0.1M HC104 at Rotating Gold Disk Electrodes." (1980). All ETDs from UAB. 7011.
https://digitalcommons.library.uab.edu/etd-collection/7011
Comments
MS - Master of Science/Master of Surgery; ProQuest publication number 31751931